Belliato M, Palo A, Pasero D, Iotti GA, Mojoli F, Braschi A.

Abstract

OBJECTIVE:

Evaluation of the respiratory pattern selected by the Adaptive Support Ventilation (ASV) in ventilated patients with acute, chronic respiratory failure and normal lungs and in a physical lung model.

DESIGN:

We tested ASV both on patients and in a physical lung model, with a normal level of minute ventilation and with minute ventilation increased by 30%. In each patient, respiratory pattern, mechanics and blood gases were recorded.

SETTING:

General ICU of a University Hospital.

RESULTS:

In patients with normal lungs, mean values+/-SD were: tidal volume (Vt) 558.1+/-142.4 mL, respiratory rate (RR) 12.6+/-1.3b/min and inspiratory time/total time ratio (Ti/Ttot) 42.4+/-4.1%; in COPD, mean values+/-SD were: Vt 724+/-171 mL, RR 9.2+/-2.7b/min and Ti/Ttot 26.6+/-10.5%; in restrictive ones, mean values+/-SD were: Vt 550.2+/-77.0 mL, RR 15.8+/-2.6b/min, Ti/Ttot 47.5+/-2.5%. In the lung model, at a normal setting, mean values+/-SD were: Vt 523+/-18.5 mL, RR 14+/-0.0b/min, Ti/Ttot 44.0%, in COPD, mean values+/-SD were: Vt 678+/-0.0 mL, RR 9+/-0.0b/min, Ti/Ttot 20+/-0.7%, in restrictive one, mean values+/-SD were: Vt 513+/-12.8 mL, RR 15+/-0.0b/min, Ti/Ttot 48+/-1.5%. In model hyperventilation conditions in a normal setting a Vt of 582+/-16.6 mL, RR 16+/-0.0b/min, Ti/Ttot 48+/-0.0% were selected, in the obstructive setting Vt 883+/-0.0 mL, RR 9+/-0.0b/min, Ti/Ttot 20+/-0.0% and in a restrictive one Vt 545+/-8.4 mL, RR 18+/-0.0b/min, Ti/Ttot 50-0.0%.

CONCLUSIONS:

In normal patients ASV selected a ventilatory pattern close to the physiological one, in COPD almost a high expiratory time pattern and in restrictive ones a reduced tidal volume pattern. In the model the selection was similar. In the hyperventilation test, ASV chose a balanced increase in both Vt and RR.

Int J Artif Organs. 2004 Aug;27(8):709-16